PDFPrintE-mail
Eurotron-Compact Combustion Analyser-GCCEurotron-Professional Combustion Analyser-GPC

Eurotron-Industrial Combustion and Emission Analyser(HH)-GHH

( Eurotron )

General

The NEW handheld EcoLine 4000 provides the latest technology in flue gas analysis. One instruments with many features:

Up to 4 gas sensors;

CO sensor with manual or automatic protection;

built-in impact printer;

water, air, surface temperature measurements;

searching for presence and location of gas leaks;

Differential pressure input for draft and gas velocity measurements;

operator safety with ambient CO continuous monitoring;

ambient temperature and relative humidity measurements.

Applications:

Boilers Utility & Power plants Industrial boilers

Heaters & Dryers Paint, Texile, Food, Paper, Rubber, etc. Process heaters

Klins / Furnaces Cement, Lime, Glass, Ceramic

Stationary internal combustion engines

Gas compression UPS power cogeneration Oil fired pumping station

Turbines

Chemical analysis laboratory Fuel additive

CEM backup & maintenance

ISO 14000 Auditing, Compliance

Features

Combustion gas analyser up to 4 gases

NO/NO & SO X 2 emissions analyser

C H un combusted X y Hydrocarbon

Draft & differential pressure meter

2 channel thermometer

Ambient CO monitor

Gas velocity with Pitot tube

Gas leak detector

Ionization flame tester

Temperature & Humidity meter

Built-in impact printer

Rechargeable battery

Industrial probe and external

compact cooler unit

Stores up to 250 samples

 


Specifications:

Type

palm-top combustion gas analyser with 1 or 2 gas sensors.

Calibration

automatic calibration procedure at instrument switch-On

Self-diagnosis

Sensors efficiency test with display diagnostic messages

Fuel types

Up to 10 selectable from keyboard.

Power supply

High capacity Ni- MH rechargeable battery pack / external battery charger

Charging time

8h at 90% with instrument Off.

Battery life

6 hours (typical) continuous use (without printing and backlight).

Memory

up to 250 full analysis data structured by boilers (Tags)

Printer

Internal impact type 24 columns with 58 mm paper roll (10 meters long).

Printer power supply

from the analyser battery pack.

Print autonomy

up to 40 reports with full battery .

Service and user information

3 programmable lines.

Printed report header

4 programmable lines.

Display

40x58 mm alpha-numeric LCD with backlight device

Flue gas probes

stainless steel shaft with incorporated temperature sensor.

Serial communication

RS232 serial interface.

Infrared port

compatible with HP82240B cordless printer.

Operating temperature

from -5°C to +45°C

Storage temperature

from -20 to +60°C (3 months maximum at temperatures exceeding the

operational limits).

Dimensions and Weight

115x90x330 mm - 1.1 kg with battery and printer

 



Parameter

Sensor

Range

Resolution

Accuracy

O2

Electrochemical

0 - 25%

0.1%

±0.2% vol

CO H2 <2000ppm comp.

Electrochemical

0 - 8000 ppm

1 ppm

±10 ppm <300 ppm

±4% up to 2000 ppm

±10% >2000 ppm

CO

Electrochemical

0 - 20000 ppm

1 ppm

±10 ppm <300 ppm

±4% up to 2000 ppm

±10% >2000 ppm

NO

Electrochemical

0 - 4000 ppm

1 ppm

±5 ppm <125 ppm

±4% up to 4000 ppm

NO2

Electrochemical

0 - 1000 ppm

1 ppm

±5 ppm <125 ppm 2

±4% up to 1000 ppm

NOx

Calculated

0 - 5000 ppm

1 ppm


SO2+9

Electrochemical

0 - 4000 ppm

1 ppm

±5 ppm <125 ppm 2

±4% up to 4000 ppm

CO2

Calculated

0 - 99.9%

0.1%


CxHy

Pellistor

0 - 5%