PDFPrintE-mail
Kimo-Laboratory Inclined Type LaboKimo-Micro manometer

Kimo-Laboratory Manometer

( Dwyer )
Feature:
  • Manufactured for the most demanding laboratory requirements, the DFP manometer has a liquid column coupled to a photoelectric measuring system that digitally displays the pressure reading, thus eliminating parallax error and ensuring accurate readings.
  • The DFP permits the calibration of sensors or other measuring equipment connected in parallel with a common pressure source.
  • The DFP is a reference standard.
Specifications:
Pressure Range0-5600 Pa; 0-9300 Pa; 0-18600 Pa dependant on the model.
Détection on 3 Levels3 LED’s to assist measurement (high, low, exact)to ensure precise alignment of the photoelectric cell with the meniscus.
Résolution±0,01 mm or 0,0005 inch.
Accuracy0,2% of the reading ± 1 Pa between 0 and 3000 Pa / ± 0,5% of the reading ± 2 Pa between 3000 and 20000 Pa.
Détection Sensitivity< to 0,03 mm.
Operating Temperature15 to 25 °C.
Operating Maximum Relative Humidity70% HR.